A Manual of Practical X-RAY Work – Chapter 2 Part 9
They should never be fully discharged in working, the safe limit being indicated by a fall in the voltage of the derived current. Each cell, as mentioned, gives a little over 2 volts when freshly charged, and the bulk of its charge (about 75 per cent.) is given off at that pressure. When the E.M.F. derived from an accumulator falls below 2 volts from each cell connected in series, it is imperative that it be recharged at once.
If, after recharging, an accumulator does not register its normal voltage, test each cell separately with a pocket galvanometer or a 2-volt lamp, so as to discover which ceil or cells are at fault.
Short of serious damage, the failure of a cell to register its full voltage is frequently due to a fall in the level of the acid, caused by leakage or evaporation. (The possibility of leakage points to the necessity of having accumulators placed on leaden trays if they be kept indoors.)
In testing for a fault, do not, as is sometimes foolishly done, spark or flash each cell by connecting its opposite plates by a piece of wire, since such short-circuiting injures the plates.
The chance of such short-circuiting by accidental means must be prevented, a possible danger of this kind in transit having been already mentioned. Thus, in connecting up the induction-coil, fasten the ends of the connecting-wires to the coil before fastening the other ends to the terminals of the accumulator, thereby avoiding the chance of live ends coming into contact. Similarly, it is well to see that the accumulator boxes are not used as a shelf for depositing odd pieces of wire or metal, which may readily bridge the terminals and cause serious damage. In extreme cases, by such short- circuiting, plates may be completely crumbled up.
With careless working, it may soon be noted that an accu- mulator will not absorb nor discharge the certified quan- tity of electricity.
Tom Thym on October 31st 2009 in x-ray
